|
Appendix K. Quality
Assurance/Quality Control for Inorganic Analyses
As noted previously, the analytical
methods employed for inorganic analyses by the environmental chemistry
program at CEMRC are based, when applicable, on various standard procedures
(EPA, 1983, Methods for Chemical Analysis of Water and Wastes, EPA/600/4-79-020;
EPA, 1997,
Test Methods for Evaluating Solid Waste, Physical/Chemical
Methods, EPA/SW-846; American Public Health Association, 1981, Standard
Methods for the Examination of Water and Wastewater, 15th Edition).
For some matrix/analyte combinations, appropriate external standard procedures
do not exist, and specialized procedures have been developed to meet the
needs of the WIPP EM.
Instrumentation
A DIONEX 500 ion chromatography (IC)
system was used to determine the concentrations of a suite of anions, including
nitrate, sulfate, chloride, fluoride, bromide, and phosphate in aqueous
extracts. Configured differently, the same instrument was used to determine
the concentrations of several cations (calcium, magnesium and potassium).
The anion analyses were performed with the use of an AS14 anion exchange
column and chemical suppression while the cations were determined using
a CG12A guard column and a CS12A analytical column.
Elemental analyses employed an atomic
absorption spectrometer (AAS) with a computer-controlled Perkin-Elmer 5100PC
atomic absorption unit with Zeeman background correction. Samples are introduced
into the AAS by aspiration through an air/acetylene flame, by vaporization
in a heated graphite furnace, by flow-injection via a heated quartz cell,
or via an unheated quartz cell (for Hg). The third instrument used for
inorganic analyses was a Perkin-Elmer 3300 dual-view, inductively-coupled
plasma atomic (or optical) emission spectrometer (ICP-ES). The AAS and
ICP-ES are complementary; AAS is more sensitive than the ICP-ES, especially
for the hydride elements (As, Sb, and Se), but compared with the ICP-ES,
the AAS has a narrower linear range, requires more operator effort for
calibration and operation, and has a much lower sample throughput.
General Quality Control
Several analytes are readily determined
by more than one of the three instruments used at CEMRC, which facilitates
intra-laboratory comparisons as summarized below. Some of these internal
QC comparisons are also summarized in the sections of this report that
deal with specific media.
Independent quality assurance samples
are obtained and analyzed to verify the performance of the instrumentation
and the proficiency of the analyst. Both blind samples (obtained from an
outside source, with true value not known at the time of analysis) and
reference samples (obtained from an outside source or prepared internally,
with true values known at the time of analysis) are used to perform this
function. Regular quality control verifications and batch QC provide records
of sample performance data. Copies of the analytical data and performance
results are maintained in the environmental chemistry instrument laboratory.
The laboratory also carried out several informal inter-laboratory comparisons,
but has not participated in a formal intercomparison program.
Calibrations are verified with a standard
obtained from a source different from that used in procuring the primary
calibration standards. The calibration standards and the verification standards
used at CEMRC are, where possible, traceable to NIST. A calibration blank
is analyzed at the beginning of each workday when samples will be run,
after every ten samples, and at the end of the day. In the calibration
verification, blank results must be less than the minimum detectable level
or + 3 SD of control limits. Analysis of a blank and a standard
are performed at a frequency of 10% during analytical runs, and these are
repeated at the close of each analysis to verify continued calibration
validity. Batch quality control samples are counted as samples in determining
the 10% frequency, but the continuing check samples are not counted as
samples in determining the 10% frequency.
Various types of field blanks, check
solutions and laboratory fortified (spiked) samples are analyzed along
with the samples as part of the QA/QC procedures. These vary somewhat among
matrices and analyses as described in more detail below. In addition, when
feasible, duplicate samples are processed to evaluate reproducibility and
sample homogeneity. Control charts for each matrix have been established,
and + 3 SD limits have been determined for future reference. Control
charts are used to track the performance of the instrument and the sample
preparation procedures. Similarly, spike recoveries are
calculated, tracked, and reported along with the analytical data.
Quality Control for Analyses
by IC
For the IC analyses, QC samples are
analyzed with each sample batch as an indicator of the reliability of the
data produced. The types, frequencies of analysis, and limits for these
QC samples have been established in a set of standard operating procedures.
Method Detection Limits (MDL) were
established for each analyte in each sample matrix according to EPA Method
300.0 (Determination of Inorganic Anions by Ion Chromatography) (Table
K1). QC samples included Laboratory Reagent Blanks (LRB), with one
LRB prepared for each sample batch (normally a set of ten samples). LRB
results below MDL are considered acceptable (Table
K2). For aerosol filter analyses, some LRB results indicated reagent
blank contamination, which was subsequently identified and eliminated.
Results for samples analyzed prior to elimination of the contamination
were corrected by subtraction of the blank value for each analyte. Laboratory
Fortified Matrix (LFM) samples were also used for QC, with one LFM analysis
per sample group. Results from analyses of LFMs are used to calculate matrix
spike recoveries, with recoveries of 70-130% considered acceptable. As
prescribed by EPA Method 300.0, chloride and sulfate values in water samples
and chloride, phosphate and sulfate values in sediments were not reported
because the concentration of the fortification was less than 25% of the
background concentration (Table
K3).
One duplicate analysis was performed
for each sample group. When feasible, duplicate aliquots of some field
samples were analyzed. In cases where duplicate aliquots from the original
sample were not feasible (such as aerosol filters), separate aliquots of
the sample extract were analyzed. The relative percent difference (RPD)
between the sample and the duplicate was calculated, with a difference
of < 20% (or an absolute difference of + MDL for samples less
than five times the MDL) considered acceptable. For aerosol filters, differences
between the chloride duplicates were not within limits when the observed
values were less than or near the MDL (Table
K4).
A Laboratory Fortified Blank (LFB)
was prepared and analyzed with each sample batch, using a spiked ultrapure
water sample for aerosol filters and water samples, and certified reference
materials (CRM) for soils and sediments. Recoveries of 85-115% were considered
acceptable for aerosol filters, sediments, and water samples. The CRM was
"Anions in Soils" from Environmental Research Associates (ERA) in Arvada,
Colorado. The preparation procedure used to certify the standard was found
to be slightly different from the procedure employed at CEMRC. ERA provided
corrected means for calculations of recoveries based on the CEMRC procedure
(Table
K5). Because there is no existing standard reference method for extracting
solid material for anion analysis by ion chromatography, the results may
not be directly comparable.
Low-volume aerosol filters were also
analyzed by IC for five cations with overall acceptable results (Table
K6). Acceptance limits for each QC parameter were the same as previously
described. In sample batches where the laboratory reagent blanks were above
the MDL, blank subtraction was performed for those affected analytes.
Quality Control for Elemental
Analyses by ICP-ES and AAS
For elemental analyses, sets of quality
control samples comparable to those previously described for IC analyses
were included with each sample batch. Detailed performance results for
all QC measures are not presented here due to the number of elements that
can be determined by ICP-ES and AAS. For all media (aerosol filters, water,
soils, and sediments), ICP-ES and AAS values were reported to the method
detection limit as determined by EPA protocols (Table
K7). Digestion QC samples were analyzed at a frequency of 10% relative
to samples. The digestion QC control parameters used for the evaluation
of metals in aerosol filters included LRB filters and vendor-supplied certified
reference filters. Due to sample volume limitations, duplicate and post
digestion spike analyses could not be performed for ICP-ES analyses of
the aerosol samples.
For water, soils, and sediments, a
practical quantitation limit (PQL) was also calculated to evaluate precision
based on the analysis of duplicate samples. The PQL is obtained by multiplying
the method detection limit (MDL) by five. The digestion quality control
parameters used for the evaluation of metals in water, soils, and sediments
were based on EPA Contract Laboratory Program (1994, U.S. EPA Contract
Laboratory Program National Functional Guidelines for Inorganic Data Review,
EPA 540/R-94013) and SW846 methods (EPA, 1997, Test Methods for Evaluating
Solid Waste, Physical/Chemical Methods, EPA/SW-846. No comparable control
parameters presently exist for aerosol samples.
For aerosol samples, unused cellulose
ester filters were used as LRB samples. LRB results above the MDL were
subtracted from each associated batch of sample results, because the LRB
results were greater than the MDL for many of the analytes studied. The
sources of this filter contamination have not yet been identified. A cellulose
ester CRM ("Trace Metals on Filter Media" from High Purity Standards in
Charleston, South Carolina) and a LFB were also used with analyses of aerosol
samples. Mean recoveries for all analytes were within 85-115% of control
limits, with the exception of Se. The CRM results for Se were 40%, 62%,
43% and 118%, while the LFB recovery for Se was 131%. The source of the
inconsistent Se recoveries is not known at this time, and therefore, Se
values for aerosol samples should be considered semi-quantitative at best.
Four standard QC measures were used
in association with analyses of water samples. Ultrapure water was used
for LRB samples and results were less than the MDLs for all analytes except
Al, Ba, Ca, Cd, Fe, Mg and K. For Al and Fe, all sample results were less
than the reagent blank, resulting in concentrations reported as <MDL.
Cd results in the samples were corrected for the LRB values in each associated
batch. For Ba, Ca, K, Mg, and K, all sample measurements were at least
10 times higher than the LRB values, and therefore the contaminant effects
for these analytes are considered negligible. A LFB was prepared by adding
a known quantity of each analyte of interest to ultrapure water. All analytes
were recovered within the 85-115% limits as specified by EPA methods. LFM
samples were also used for QC in analyses of water samples, with all recoveries
within the 85%-115% acceptance window specified in the EPA methods, with
the exception of Pb. Although Pb was recovered at 81%, this level is considered
usable according to EPA inorganic usability protocols. A duplicate digestion
analysis of water samples was also performed to demonstrate reproducibility,
but a slight modification of the EPA CLP program was used for acceptance
determination. If the sample result was less than the PQL, a + PQL
control limit was used. If the sample result was greater than the PQL,
a +20% RPD control limit was used. All duplicate results were within
these modified acceptance limits.
For soils and sediments, LRB samples
of ultrapure water were compared to MDLs to determine if contamination
was introduced during sample preparation. The LRB results were above the
MDLs for Ca, Mg, and Fe. However, the sample measurements were several
orders of magnitude higher than the LRB results for these analytes, and
therefore the contaminant effects on the measurements were considered negligible.
If the LRB result were greater than the MDL, a correction to the sample
results was made. Several of the LRB results for Cu, Cd, Ni and Na were
at approximately the same level as those measured in the soil samples (within
a factor of five), and therefore, results for these analytes may be biased
high. These results would be considered "estimated" but would be considered
usable according to EPA inorganic data usability protocols (1994, U.S.
EPA Contract Laboratory Program National Functional Guidelines for Inorganic
Data Review, EPA 540/R-94013.) All other LRB results were within acceptance
limits for soils. The elemental concentrations of all analytes in sediment
samples were at least ten times higher than LRB results, and therefore
the contaminant effects on measurements in sediments are considered negligible.
A CRM ("Priority Pollutant T/CLP Soil"
from ERA) was obtained and prepared with the soil and sediment samples
to demonstrate matrix-specific performance of digestion and analysis procedures.
All analytes were recovered within the supplier’s specified control limits
for all digestions. The average CRM recoveries were within 85%-115% for
all analytes, with the exception of Zn at 73% and Sb at 56%. A low bias
for Sb was expected, due to use of a standard hotplate digestion procedure
that allows loss of Sb as a SbCl precipitate. Additional studies are underway
to resolve the low bias for Zn. Duplicate digestions were preformed
for soil and sediment using a modification of the EPA CLP program for acceptance
determination. If the sample result was less than the PQL, a + PQL
control limit was used. If the sample result was greater than the PQL a
+20%
RPD control limit was used. For soils, the average RPD over the nine digestions
performed was within acceptance limits for all analytes.
Al, As, Ba, Ca,
Cr, Fe, Mg, Ni, K, V, and Zn each had one of the nine digestions outside
the acceptance limits. None of the RPDs were outside the limits established
in the EPA usability protocols. For sediments, the duplicate precision
for Mo in the first digestion was 22%, which is outside the acceptance
limit. However, the Mo data are considered usable according to the EPA
inorganic usability protocols. All other duplicate results for the sediments
were within acceptance limits. A LFM also was prepared for soil, with an
average recovery for the nine digestions within 85%-115% windows for all
analytes with the exception of Sb at 37%. All individual recoveries were
within the 70%-130% acceptance window specified in the EPA methods with
the exception of Sb and Se in one sediment digestion. As previously noted
a low bias for Sb was expected due to the digestion procedure used. Experiments
to resolve the bias for Se are in progress.
|